User’s Guide
MultiMaster™ Digital MultiMeter Series
With PC Interface
Models:
MM560
MM570
Vue de la page 0
1 2 3 4 5 6 ... 20 21

Résumé du contenu

Page 1

User’s Guide MultiMaster™ Digital MultiMeter Series With PC Interface Models: MM560 MM570

Page 2 - P/N: 7M1C-0271-A000

9Auto Power Off: After 17 minutes of inactivity with no input signal. Power consumption during auto power off; 20µA Safety: The MultiMaster™ Series

Page 3

10Meter Description (MM570 Shown) 1. Liquid Crystal display 2. Function push-keys 3. Rotary function switch 4. 10A/-T2 input jack 5. V/Hz/Ω/Cap/+T1

Page 4

11DC+AC True RMS DC+AC True RMS calculates both of the AC and DC components given by the expression when making measurement, and can responds accura

Page 5 - Specifications

12Analog bar-graph The analog bar graph provides a visual indication of measurement like a traditional analog meter needle. It is excellent in detect

Page 6

13Operation Measurement Considerations NOTICE: Read and understand all warning and caution statements listed in the safety section of this operation

Page 7

142. For current measurements up to 5000µA, set the function switch to the "µA" position and insert the red test lead into the mA-µA jack

Page 8

156. For Continuity tests, if the resistance is less than the threshold (20Ω to 200Ω), an audible tone will sound.

Page 9

16Capacitance Measurements 1. Insert the black lead into the negative COM jack and the red test lead into the positive CAP jack. 2. Set the function

Page 10

17Diode Test 1. Insert the black lead into the negative COM jack and the red test lead into the positive CAP jack 2. Set the function switch to the &

Page 11

18 Relative mode Relative Zero allows the user to offset the meter by using the displayed value as the zero reference value. Practically all of the d

Page 12

1WARRANTY EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for three years from date of shipmen

Page 13

19RS-232 PC Interface This instrument is equipped with an optically isolated interface port (located on the rear of the meter). An optional WindowsTM

Page 14

20 Support Hotline (781) 890-7440 Tech support: Ext. 200; Email: [email protected] Repair/Returns: Ext. 210; Email: [email protected] Website: www.

Page 15

2Introduction Congratulations on your purchase of Extech model MM560 or MM570 digital multimeter. Properly used, this meter will provide many years o

Page 16

37. Use great care when making measurements if the voltages are greater than 25VAC rms or 35VDC. These voltages are considered a shock hazard. 8. A

Page 17

4Specifications DC VOLTAGE Accuracy Range MM570 MM560 500.00 mV, 5.0000V, 50.000V 0.02% + 2d 0.03% + 2d 500.00V 0.04% + 2d 0.05% + 2d 1000.0V 0.

Page 18

5AC & AC+DC VOLTAGE Accuracy* MM570 MM560 Range 20Hz to 45Hz 500.00mV, 5.0000V, 50.000V 1.5% + 40d 500.00V, 1000.0V unspecified unspecified

Page 19

6AC & AC+DC CURRENT Accuracy MM570 MM560 Range 50Hz to 60Hz Burden Voltage 500.00µA, 5000.0µA 0.15mV/µA 50.000mA, 500.00mA 3.3mV/mA 5.0000A,

Page 20

7CAPACITANCE Range Accuracy* 50.00nF, 500.0nF 0.8% + 3d 5.000µF 1.0% + 3d 50.00µF 2.0% + 3d 500.0µF 3.5% + 5d 9999µF 5.0% + 5d *Accuracies with fil

Page 21

8DC Loop Current %4 to 20mA: 4mA = 0% (zero), 20mA = 100% (span), Resolution: 0.01%, Accuracy: ±25d Diode Test: Accuracy: 1%+1d, Current 0.8mA(typic